The Human Testosterone CLIA Kit is a Chemiluminescence Immunoassay (CLIA) intended for the quantitative measurement of human testosterone concentration in serum.
Compatible with the ECL100 Chemiluminescent Immunoassay Analyzer.
For research use only. Not for use in diagnostic procedures.
For lab professional use only.
Manufactured by Epitope Diagnostics, Inc.
Please contact us for a quote.
The Testosterone CLIA Kit is designed, developed, and produced for the quantitative measurement of testosterone level in serum samples. The assay utilizes competitive bonding mechanisms for testing.
The calibrator, control or sample, and Biotinylated Testosterone monoclonal antibody are mixed and incubated to form antigen-antibody immune complex.
Streptavidin coated particles and Acridinium ester Testosterone are added under culture conditions. The unbound biotinylated Testosterone monoclonal antibody of the sample binds to Acridinium ester Testosterone and binds to the microparticle through the biotin-streptavidin reaction. After precipitation in a magnetic field, the supernatant is decanted, and then a wash cycle is performed to remove any remaining substances that are not bound to the magnetic microbeads.
Subsequently, the washed compound is sent into the measurement chamber where trigger solution is automatically added to initiate a chemiluminescence reaction. The light signal is measured by a photomultiplier as relative light units (RLUs), which is inversely proportional to the concentration of Testosterone present in the sample.
The test result is automatically calculated by the system according to the working curve.